Autoclave water: why distilled water, and the purity limits by brand
Use distilled water for an autoclave — every tabletop manufacturer says so, and most give a number. SciCan wants under 10 µS/cm in a Statim; Tuttnauer and the Bravo allow up to 15. Water is the most common root cause we find behind sensor, valve and boiler faults, and the cheapest one to prevent.

The limits, by brand
Conductivity, in microsiemens per centimeter (µS/cm), is the one number every manual gives. Lower is purer.
| Unit | Water | Conductivity | Other limits |
|---|---|---|---|
| SciCan Statim 2000 / 5000 | Steam-process distilled | under 10 µS/cm | 5 ppm or less dissolved solids |
| Tuttnauer EZ9Plus / EZ11Plus | Distilled or mineral-free | ≤ 15 µS/cm | pH 5–7.5, hardness ≤ 0.02 mmol/L, chloride ≤ 2 mg/L (EN 13060 table) |
| SciCan Bravo G4 | Demineralized or distilled | under 15 µS/cm | pH 5–7, hardness under 0.02 mmol/L |
| Midmark M9 / M11 | Distilled, or water meeting the user-guide purity spec | under 10 µS/cm* | pH 5–7.5*; older sheets cite AAMI ST46: evaporated residue ≤ 15 mg/L, silica ≤ 2 mg/L |
* The Midmark figure comes from the AAMI ST108 reference table in Midmark's Auto Fill Kit guide for the M9 and M11. We have not yet confirmed it against the full PDF, so treat it as to be confirmed and check the guide for your unit. Plain distilled water meets it comfortably.
Tuttnauer also gives a limit for the condensate — the steam that comes off an empty chamber — of 3 µS/cm, and recommends testing your water monthly. It notes that water outside the table can invalidate the guarantee.
What tap water does inside the unit
Tap water carries dissolved minerals and chlorides. When it boils in a chamber or a steam generator, the water leaves and the minerals stay. Over weeks that shows up in specific parts, and the manuals' own fault tables point at them:
- Water-level sensors and electrodes. Scale and residue on the probe make the unit misread the water level. Tuttnauer lists a dirty or shorted water electrode as a cause of LOW WATER and tells users to clean the sensor's sides weekly; Midmark lists a dirty water-level sensor among its fill-fault causes.
- Conductivity sensors. The Statim and the Bravo measure water quality themselves. The Statim stops with a "water quality unacceptable" message above its 10 µS/cm default threshold; the Bravo measures conductivity at the start of every cycle, prints it on the report and, with poor water, allows five more cycles before it locks until the water is changed.
- Valves. Fill, vent and solenoid valves seat against small orifices. Mineral grit keeps them from closing fully — the kind of leak that later shows up as a pressure or temperature fault.
- Filters, pumps and lines. Clogged screens and pump lines starve the unit of water. On a Statim, SciCan notes that poor water delivery is a frequent cause of boiler overheat, and that a blown thermal fuse is the symptom, not the cause.
- The chamber and instruments. Midmark names poor water purity among the causes of rust on the M9 and M11. Chloride is also on every manufacturer's limit list for a reason: it attacks stainless steel.
We see the result on the bench as scaled heaters, sensors that need cleaning before any diagnosis makes sense, and valves that stick. None of it shows up on the first day; all of it shows up eventually.
How to check your water
- Buy steam-distilled water and read the label. SciCan's Bravo manual tells you to make sure the declared characteristics match its table.
- Measure it. SciCan sells a conductivity meter (part 01-103139S) and accepts any equivalent; inexpensive handheld meters are fine for this. Test at room temperature — SciCan points out that temperature has a large effect on the reading.
- Use the unit's own test if it has one. The Bravo has an H2O test option and shows the value on each cycle report.
- Keep the reservoir fresh. Drain and refill weekly on Midmark and Tuttnauer units, as their schedules ask. Water that sits picks up residue from the chamber.
Distilled, deionized, reverse osmosis
The manuals specify the water by its numbers, not by how it was made: Tuttnauer says "distilled or mineral-free," SciCan says "demineralized or distilled" for the Bravo and steam-process distilled for the Statim. Water that meets the table is acceptable; water that does not, whatever the label says, is not. If your office runs an in-house water system, test its output against your unit's limit rather than assuming it passes.
Things not to put in the reservoir
- Tap water or filtered drinking water.
- Rust inhibitors or instrument milk — Midmark lists them as a cause of odor.
- Descalers the manufacturer did not name. SciCan specifically warns against using CLR on a Statim boiler.
- Chlorine or bleach of any kind.
If the damage is already done
Switching to distilled water stops new scale, but it does not undo what is already on a sensor or in a valve. If a unit that ran on tap water now throws fill, low-water or water-quality faults even on fresh distilled water, it needs cleaning inside — and sometimes a valve or sensor. Our bench maintenance ($349) covers the cleaning, filters and gaskets; a stuck valve is quoted from $399 on the price list. For the cleaning you can do yourself, see how to clean an autoclave, and for specific messages, Statim cycle faults and Tuttnauer error messages.
Questions
Can I use tap water in an autoclave?
No. Every tabletop manufacturer we cover — Midmark, Tuttnauer and SciCan — specifies distilled or demineralized water. Tap water leaves minerals and chlorides in the chamber, on the water-level and conductivity sensors and in the valves, and several units will refuse to run on it.
Is distilled water from the grocery store good enough?
Often, but not always. What matters is that it meets your unit’s limit, for example below 10 µS/cm for a Statim or 15 µS/cm for a Tuttnauer EZPlus or Bravo. SciCan’s Bravo manual tells you to check that the declared characteristics on the bottle match its table. A cheap conductivity meter settles the question.
What is µS/cm?
Microsiemens per centimeter, a measure of how well water conducts electricity. Pure water conducts very little; dissolved minerals and salts raise the number. It is the simplest single check of water purity, which is why autoclave manuals use it.
My Statim says water quality unacceptable. Is it broken?
Usually not. On a Statim the default bad-water threshold is 10 µS/cm, and the message means the reading is above it. Drain the reservoir, refill with fresh steam-distilled water at room temperature and try again. SciCan says not to raise the threshold to suit the water, because that can damage the unit.
Still not working?
Then the fault is inside the unit. We pick it up anywhere in Southern California, find the fault on the bench and quote it before any work — $125 for pickup, diagnostic and return, credited toward the repair. A loaner can ride along on the same stop.
Sources
- SciCan Statim 5000 / 5000S / G4 service manual — water conductivity testing and setup