Statim 2000 repair · Southern California

Statim 2000 repair: Classic, 2000S and Statim 2000 G4

The Statim 2000 is the small-cassette Statim — the fast chairside sterilizer for handpieces and a few instruments between patients. We repair the Statim 2000, 2000S and 2000 G4 on the bench, including the oldest units that still show Check Cassette messages, and prove every repair on full cycles.

Models we see: Statim 2000 Classic (including revision 3.x/4.x Check Cassette units), Statim 2000S, Statim 2000 G4.

Cassette-style tabletop sterilizer with its stainless cassette pulled out, a replacement cassette seal beside it

What the Statim 2000 is

The Statim 2000 is the smaller of SciCan's two cassette sterilizers. A pump meters distilled water into a small steam generator, the steam goes straight into the cassette, and a solenoid valve opens and closes the exhaust. The cassette is the chamber, and the seal in it does the job a door gasket does on a chamber autoclave. The principle, and why it fails the way it does, is on our SciCan Statim page.

Three things about the 2000 are specific to it:

  • The venturi in the cassette tray. The 2000 cassette has a venturi at the left rear of the tray. When it clogs, the cassette does not vent properly: you get CF7, CF8 or CF10, and often a cassette that is hard to pull out after the cycle. Cleaning it is something the office can do.
  • Its own filters. The compressor draws air through a round foam filter behind a plate on the rear of the compressor — reachable without taking the cover off — and then through the bacteria-retentive filter on the rear cover. Change both at the same time.
  • The oldest boards speak a different language. Statim 2000 non-S units with revision 3.x or 4.x controllers show "Check Cassette" and a number, and no software revision at power-up. Every later 2000 shows "CF" numbers from the shared cycle fault list.

What it does, what failed, what it costs

SymptomUsual causeWhat we doPrice
Cycle fault 3 or 4 (or Check Cassette 3 on old units), steam at the cassetteCassette seal past 500 cycles or 6 months, warped or damaged cassette, steam tubing leak, solenoid stuck openNew seal, cassette checked, steam circuit leak-tested, full cycles loggedfrom $249
CF7 and the cassette is hard to pull out after ventingClogged venturi at the left rear of the cassette tray, kinked exhaust tube, solenoid not openingVenturi and exhaust cleared, solenoid plunger cleaned or the valve rebuiltfrom $399
CF10, CF11, CF15 — will not purge, vent or coolKinked or pinched exhaust tube, clogged venturi, solenoid stuck closed or a failed coilExhaust path cleared, solenoid rebuilt or replacedfrom $399
CF16, CF20, CF27 — steam generator overheatWeak pump, blocked pump filters, kinked pump tube; G4: float switch stuck upPump flow tested, filters and tube replaced, pump replaced if weakfrom $499
CF1, CF25 or Check Cassette 1 — never heatsBlown thermal fuse after an overheat, open steam generator, no drive from the boardFuse, generator and board measured; the water-delivery fault behind it fixedfrom $449
Wet wraps, noise in drying, GFI trips at power-upDirty compressor foam filter or bacteria filter, leaking check valveFilters replaced, check valve or compressor replacedfrom $399
G4: CF94, CF95, CF96, blank or frozen color screenRevision 7 board memory, display controller or the cable between themBoard, cable or display controller repaired or replaced, software restoredfrom $449

Prices are "from", with labor, ordinary parts and a test cycle. The final price comes after the bench diagnostic, and nothing is done until you approve it. All prices →

The Statim 2000 G4

The 2000 G4 keeps the same cassette, pump, steam generator and solenoid design, so its mechanical faults are the ones above — cycle fault 3 from a worn seal, overheat faults from water delivery, exhaust faults from a clogged venturi or a sticking solenoid. The G4 changes are electronic:

  • A revision 7 controller with a built-in service menu and NVRAM tools — the unit can be diagnosed with the cover on, and a lost chamber calibration (CF90) is first tried with an NVRAM restore.
  • A color display with its own controller, and its own faults: CF95 (the display stops hearing from the main board) and CF96 (the display's program fails its check at power-up).
  • A float switch in the reservoir that can stick in the raised position and cause CF16 or CF27 with an empty reservoir.
  • No STAT-DRI on the US G4 models, per SciCan.

Statim 2000 maintenance

  • Daily: steam-process distilled water only (under about 10 µS/cm); empty the condenser bottle at each refill and keep it at the MIN line.
  • Weekly: wash the cassette with a non-chlorine detergent and check the filters. Clean the venturi whenever the cassette starts venting slowly or sticks.
  • Every 500 cycles or 6 months: new cassette seal, new foam and bacteria-retentive filters.

Our bench preventive maintenance covers all of it for $349, with a test cycle and the printout. Other repairs are on the pricing page, and a Statim loaner can ride along on the pickup.

What the price includes

  • Pickup and return anywhere in our Southern California area, the bench diagnostic and a written quote — $125, credited toward the repair.
  • After the repair: full cycles on the bench with a temperature and pressure check, the printout or log kept for your records.
  • A loaner on the same stop if you need one — from $149, free on repairs over $900.
  • 90-day warranty on repairs, 1 year on overhauls.

Questions

What is the difference between the Statim 2000 and the Statim 5000?

Mostly the cassette. The 2000 takes a smaller cassette and is usually a chairside or second unit; the 5000 is the large-cassette main sterilizer. The working parts — pump, steam generator, solenoid valve, check valve, compressor — follow the same design, and both share the CF cycle fault list. The 2000 has its own cassette with a venturi in the tray, and its own filter arrangement.

Our old Statim 2000 says "Check Cassette" with a number. Is that the same as a cycle fault?

Not exactly. "Check Cassette" plus a number comes from the oldest Statim 2000 boards (revision 3.x and 4.x), which show no software revision at power-up. Their numbers have their own meanings — Check Cassette 1 to 3 are close to CF1 to CF3, but the higher numbers do not line up with the CF list. Tell us the exact message and we will read it against the right chart.

How often should the Statim 2000 cassette seal be replaced?

Every 500 cycles or 6 months, whichever comes first, per SciCan, together with the air filters. A 2000 seal is roughly $75–95 as a part; we fit it from $249 with a test cycle.

Is there a loaner while our Statim 2000 is repaired?

Our Statim loaners are Statim 5000s with their own cassette — $199 for up to 21 days, free on repairs over $900. It is the bigger cassette, so the instruments fit and your cycles stay the same kind of fast cassette cycles.

Repair or replace a Statim 2000?

Dealers list a new Statim 2000 G4 anywhere from about $5,000 to $7,100. Most 2000 repairs are a seal, a solenoid, a pump or a board — $249 to $899 — and a full overhaul is $1,595. On the very oldest Check Cassette units, parts availability is the question, and we tell you before you pay for more than the diagnostic.

Southern California pickup

Book a pickup

Tell us the model, the error on the screen and what it does. A technician reads it and texts you back with a price range and the next pickup day in your area.

  • Pickup, bench diagnostic and return — $125, credited toward the repair
  • A tested loaner swapped in on the same stop, from $149 — free on repairs over $900
  • A written quote before any work starts
  • Typical bench time: 3 to 5 business days on the bench, parts permitting