Tuttnauer 3870 repair · Southern California

Tuttnauer 3870 repair

The Tuttnauer 3870 is the largest tabletop Tuttnauer: an 85-liter chamber for surgical packs, veterinary kits and the busiest dental practices. It fails the same way the smaller Tuttnauers do — water, seals, elements — but its size and its 230-volt supply change how the faults show up.

Models we see: Tuttnauer 3870M (manual); 3870E and 3870EAP (automatic).

Open chamber of a tabletop autoclave with trays removed, cleaning brush and removed drain screen in front

The 3870 in numbers

Tuttnauer lists the 3870M with a 15 × 30 inch chamber of 85 liters, two long trays and room for 15 full-size cassettes. It runs a standard unwrapped cycle in about 23 minutes from a hot start and 32 from cold, and it switches itself off at the end of the sterilize and dry phases. The 3870E and 3870EAP automatics put a microprocessor on the same chamber size and report faults with the same text messages as the other E-series Tuttnauers. (Some sources give the chamber as 15 × 28 inches; go by the label on your unit.)

The numbers that matter for repair are the electrical ones: 230 V, 3000 W, 13.6 A. A 3870 on an undersized or low-voltage circuit heats slowly enough to throw LOW HEAT on an automatic model or just drag on a manual one, with nothing wrong inside. And with that much chamber to heat, a weak element shows up as a unit that still works, only slowly.

What it does, what failed, what it costs

SymptomUsual causeWhat we doPrice
Slow heat-up, LOW HEAT on the automatic modelsA failed heating element, a supply below the 220–235 V the unit needs, or a safety thermostat opening earlyEach element and the supply measured under load, element or thermostat replacedfrom $449
Steam at the door, LOW TEMP or LOW PRESThe long door gasket on the 15-inch door wears and flattens, or the door no longer closes squareGasket replaced, door hinge and locking pins adjusted, full-load test cyclefrom $249
LOW WATER or the chamber boils dryScaled water electrode or fill line, a leaking valve, or the unit not level on a long chamberElectrode and fill path cleaned, leaks traced, unit leveledfrom $399
HI TEMP or HI PRESShorted solid-state relay or element, faulty sensorRelay, element and sensor tested and replacedfrom $449
Wet packs on full surgical loadsOverloaded chamber, scale in the exhaust path, or weak drying heatChamber descaled, exhaust cleaned, drying checked on a full wrapped loadfrom $349

Prices are "from", with labor, ordinary parts and a test cycle. The final price comes after the bench diagnostic, and nothing is done until you approve it. All prices →

What you can check before calling

  • Unplug the unit and let it cool completely. Never open the door or drain while there is pressure in the chamber.
  • Make sure it is level side to side and front to back — a long chamber that tilts fills unevenly and can report LOW WATER.
  • Drain the reservoir and refill with distilled water; clean the chamber water electrode on automatic models.
  • Wipe the door gasket and look for flattened or cracked spots along its full length.
  • Have an electrician confirm the outlet voltage under load if heat-up times have crept up across the board.

Then look the message up on our Tuttnauer error messages page. If it keeps coming back, the unit needs the bench. Heating elements start at $449, gaskets at $249, and an automatic overhaul is $1,595 — see the full price list. For smaller manual units, the 2540M page covers the same failure modes.

What the price includes

  • Pickup and return anywhere in our Southern California area, the bench diagnostic and a written quote — $125, credited toward the repair.
  • After the repair: full cycles on the bench with a temperature and pressure check, the printout or log kept for your records.
  • A loaner on the same stop if you need one — from $149, free on repairs over $900.
  • 90-day warranty on repairs, 1 year on overhauls.

Questions

What electrical supply does a Tuttnauer 3870 need?

Tuttnauer lists the 3870M at 230 V, 3000 W and 13.6 A, needing 220–235 V, and notes that a 0.5 kVA buck/boost transformer may be needed. On a supply below that range the chamber heats slowly, and that can look like an element fault. We check the supply before we condemn a part.

Can a courier carry a 3870?

Yes, with two people. Tuttnauer gives the 3870M shipping weight as 278 lb including the stand. Tell us on the form that it is a 3870 so two people come for it.

Do you have a loaner for the 3870?

Our loaner fleet is M9 and M11 class units with much smaller chambers. One can keep instruments moving while the 3870 is out, but it will not take a full surgical pack load. Ask when you book and we will say honestly whether it helps your workflow.

Southern California pickup

Book a pickup

Tell us the model, the error on the screen and what it does. A technician reads it and texts you back with a price range and the next pickup day in your area.

  • Pickup, bench diagnostic and return — $125, credited toward the repair
  • A tested loaner swapped in on the same stop, from $149 — free on repairs over $900
  • A written quote before any work starts
  • Typical bench time: 3 to 5 business days on the bench, parts permitting

Rather talk? Call or text (424) 622-8416.