Statim cycle fault codes: the full CF list and the screen messages
Every Statim cycle fault — CF1 to CF30, the cycle checks in the 50s and 60s, and the board faults from CF70 to CF99 — in plain words, from SciCan's service manual. For each one: what it means, what usually causes it, and what you can check at the office before the unit goes to the bench.

Before you look up a code
Statims show faults in one of three ways, depending on age:
- "Check Cassette" plus a number — only the oldest Statim 2000 non-S units with revision 3.x/4.x boards. They show no software revision at power-up. Check Cassette 1 to 3 are close to CF1 to CF3 below (no warm-up to 95 °C; no climb to 100 °C; no pressurization to 110 °C), but the higher numbers follow a different chart.
- "CF" plus a number — every Statim 2000 and 5000 with revision 2.x, 5.x, 6.x or 7.x boards, G4 included. This page covers these.
- A message with no number — process problems rather than failures, listed further down.
Write down three things: the model (2000 or 5000, S or not), the software revision that appears at the top right of the screen for a few seconds after power-up (for example S2S2R415 — the "R415" is what matters), and the CF number. The causes below are SciCan's guide, not a verdict: the code is the first problem the software noticed, and SciCan itself warns that a blown thermal fuse, for example, is a symptom of an overheat whose cause is often poor water delivery.
What you can check at the office
Most codes in the first group can be narrowed down without tools:
- The seal and the cassette. SciCan wants a new cassette seal every 500 cycles or 6 months. Look for a flattened, cracked or hard seal, debris in the seal channel, and a lid or tray that is dented or does not sit flat.
- The exhaust path. The exhaust tube at the back must have no kinks and be pushed fully into its fitting. On a Statim 2000, the venturi at the left rear of the cassette tray clogs.
- Water. Steam-process distilled water only, reservoir filled, condenser bottle at its MIN line. On a G4, an empty reservoir with the float stuck up causes overheat faults.
- The load. An overloaded cassette causes warm-up faults (CF1, CF2).
- The unit is level, and the cassette goes fully home (you should hear the solenoid click when you insert and remove it).
Process faults: CF1 to CF30
Numbers 9, 13, 17–18, 21–24 and 29 are not used.
| Code | What it means | Likely causes | What to do |
|---|---|---|---|
| CF1 | The cassette did not reach 95 °C in the time allowed (warm-up). | A very large cassette leak or an overloaded cassette; poor pump delivery; blown thermal fuses; contaminated or open steam generator; failed generator triac on the board. | Run a lighter load and check the seal. If it repeats, the steam generator and its fuses need measuring — a blown fuse means it overheated, so the pump is checked too. |
| CF2 | The cassette did not climb from 95 °C to 100 °C in the time allowed. | Major steam leak, overloaded cassette, chamber calibration off. | Look at the cassette seal (past 500 cycles or 6 months?) and the cassette for warping or dents; replace the seal if in doubt. Lighten the load. If it repeats: calibration check on the bench. |
| CF3 | The cassette did not pressurize and reach 110 °C in the time allowed. The most common Statim fault. | Cassette seal, damaged cassette, steam tubing leak, thermocouple probe misaligned, solenoid stuck open, check valve or pressure relief valve leaking. | Look at the cassette seal (past 500 cycles or 6 months?) and the cassette for warping or dents; replace the seal if in doubt. Full walkthrough on the cycle fault 3 page. |
| CF4 | Sterilization conditions were not reached in the time allowed. | Same list as CF3: seal, cassette, tubing leak, probe alignment, solenoid stuck open, check valve, PRV. | Look at the cassette seal (past 500 cycles or 6 months?) and the cassette for warping or dents; replace the seal if in doubt. Then treat it as CF3. |
| CF5 | Older software (1.xx) only: the pump was asked for water outside its expected window. | Leaking cassette, kinked pump tube, leaking solenoid valve, poor pump delivery, overloaded cassette. | Look at the cassette seal (past 500 cycles or 6 months?) and the cassette for warping or dents; replace the seal if in doubt; lighten the load. Otherwise a bench check of pump and solenoid. |
| CF6 | A thermocouple (steam generator on 2.xx software, validation thermocouple on later) reads 5–6 °C above the cassette during sterilization. | Seal misaligned or debris in the seal channel, cassette steam leak, sticking solenoid, kinked exhaust tubing, calibration imbalance, contaminated steam generator. | Look at the cassette seal (past 500 cycles or 6 months?) and the cassette for warping or dents; replace the seal if in doubt; clean the seal channel. Check the exhaust tube at the back for kinks and that it is pushed fully into its fitting; on a 2000, clean the venturi at the left rear of the cassette tray. If it repeats: recalibration on the bench. |
| CF7 | Cassette temperature dropped below its threshold during sterilization. | Solenoid stuck open, cassette leak, check valve or PRV leak. On a 2000 with a cassette hard to remove after venting: kinked exhaust, clogged venturi, solenoid not opening. | Look at the cassette seal (past 500 cycles or 6 months?) and the cassette for warping or dents; replace the seal if in doubt. Check the exhaust tube at the back for kinks and that it is pushed fully into its fitting; on a 2000, clean the venturi at the left rear of the cassette tray. |
| CF8 | A thermocouple reads 5–6 °C below the cassette during sterilization. | Contaminated or failed steam generator, calibration imbalance, high pump flow, blocked cassette exhaust duct, sticking solenoid. | Check the exhaust tube at the back for kinks and that it is pushed fully into its fitting; on a 2000, clean the venturi at the left rear of the cassette tray. If it repeats: solenoid check and recalibration on the bench. |
| CF10 | During conditioning (air purge) the cassette did not drop to 115 °C (110 °C on Rubber and Plastics). | Blocked cassette exhaust duct, kinked exhaust tube, solenoid stuck closed, failed solenoid coil. | Check the exhaust tube at the back for kinks and that it is pushed fully into its fitting; on a 2000, clean the venturi at the left rear of the cassette tray. Listen for the solenoid click when you insert and remove the cassette — no click points to the valve. If the exhaust runs to a VistaCool or StatFlow system, check its water supply is on. |
| CF11 | After the cycle, the cassette did not cool to 102 °C during venting in time. | Same as CF10: exhaust duct, kinked tube, solenoid stuck closed or failed coil. | Check the exhaust tube at the back for kinks and that it is pushed fully into its fitting; on a 2000, clean the venturi at the left rear of the cassette tray. |
| CF12 | A fault in the temperature measuring system. | A thermocouple (they read about 10 ohms at room temperature) or the board. | Bench repair |
| CF14 | Steam generator above 171 °C during sterilization. | Poor pump delivery, calibration imbalance, failed water level sensor. | Fill the reservoir with distilled water; on a G4 check the float in the reservoir is not stuck up. Then it needs a pump flow test on the bench |
| CF15 | Cassette temperature above its high limit during sterilization, or above 138.6 °C during conditioning or pressurizing. | Blocked exhaust duct, kinked exhaust tube, solenoid stuck closed or failed coil. | Check the exhaust tube at the back for kinks and that it is pushed fully into its fitting; on a 2000, clean the venturi at the left rear of the cassette tray. |
| CF16 | Steam generator overheat. | Blocked pump filters, weak or failed pump, kinked pump tube, failed pump or generator triac, solenoid stuck closed, seal installed wrong. G4: reservoir float switch stuck in the raised position. | Fill the reservoir with distilled water; on a G4 check the float in the reservoir is not stuck up. Then it needs a pump flow test on the bench. Check the seal is seated correctly. |
| CF19 | Calibration invalid — the steam generator calibration (software R1xx, R2xx, R4xx) or the validation thermocouple calibration (R5xx, R6xx). | A new controller board or microprocessor was fitted, or a strong static discharge corrupted memory. | Bench: recalibration; if it keeps returning, the microprocessor and EEPROM (a matched set) are replaced. |
| CF20 | The pump failed to cool the steam generator: still above 140 °C for 3.6 seconds after pumping. | Blocked pump filters, weak or failed pump, kinked pump tube, triacs, solenoid stuck closed, seal installed wrong, failed water level sensor. | Fill the reservoir with distilled water; on a G4 check the float in the reservoir is not stuck up. Then it needs a pump flow test on the bench |
| CF25 | No call for water within 90 seconds of the start of the cycle. | The steam generator is not heating: blown thermal fuse (check pump delivery and the triac too), open steam generator. | Bench repair. The fuse is a symptom of an earlier overheat; the water delivery is checked before it is replaced. |
| CF26 | The sterilization phase did not start within 3 minutes of reaching temperature, three cycles in a row (the first two show "Cycle Interrupted"). | Calibration imbalance between the thermocouples, poor pump delivery, sticking solenoid valve, or a reference voltage (Vref) out of spec on the board. | Look at the cassette seal (past 500 cycles or 6 months?) and the cassette for warping or dents; replace the seal if in doubt. If it repeats: solenoid and calibration check on the bench. |
| CF27 | The steam generator did not cool below its set point (150 °C or 165 °C) in time. | Poor pump delivery, failed water level sensor, contaminated steam generator. G4: float switch stuck up. | Fill the reservoir with distilled water; on a G4 check the float in the reservoir is not stuck up. Then it needs a pump flow test on the bench |
| CF28 | S models: cassette pressure above its ceiling. | Pressure measurement failure, calibration imbalance between pressure transducer and thermocouple, solenoid stuck closed or failed coil, blocked exhaust, restricted transducer tube. | Check the exhaust tube at the back for kinks and that it is pushed fully into its fitting; on a 2000, clean the venturi at the left rear of the cassette tray. Otherwise bench. |
| CF30 | S models: the measured temperature lagged the temperature expected from the measured pressure by more than 15 seconds. | Poor air removal during conditioning (exhaust blockage), calibration imbalance, restricted pressure transducer tube. | Check the exhaust tube at the back for kinks and that it is pushed fully into its fitting; on a 2000, clean the venturi at the left rear of the cassette tray. Otherwise bench. |
Cycle limit checks: CF50 to CF65
These are SciCan's checks during the sterilization hold itself: CF50–55 for the 121 °C Rubber and Plastics cycle, CF60–65 for 134 °C cycles. The ones that compare temperature with pressure use the pressure transducer, so they belong to the S models. They all trace back to the same few things: air not purged from the cassette, a leak, the solenoid, a kinked exhaust or calibration. Numbers 31–49 are not used.
| Code | What it means | Likely causes | What to do |
|---|---|---|---|
| CF50 | Rubber and Plastics (121 °C): cassette temperature fell below 121 °C. | Poor air removal or blocked exhaust, solenoid stuck closed or failed coil, kinked exhaust tubing, calibration imbalance between the pressure transducer and the chamber thermocouple; for the low readings also a steam leak or not enough steam. | Look at the cassette seal (past 500 cycles or 6 months?) and the cassette for warping or dents; replace the seal if in doubt. Check the exhaust tube at the back for kinks and that it is pushed fully into its fitting; on a 2000, clean the venturi at the left rear of the cassette tray. |
| CF51 | Rubber and Plastics: cassette temperature rose above 125 °C. | Poor air removal or blocked exhaust, solenoid stuck closed or failed coil, kinked exhaust tubing, calibration imbalance between the pressure transducer and the chamber thermocouple; for the low readings also a steam leak or not enough steam. | Check the exhaust tube at the back for kinks and that it is pushed fully into its fitting; on a 2000, clean the venturi at the left rear of the cassette tray. |
| CF52 | Rubber and Plastics: the temperature worked out from pressure is more than 2 °C below the measured temperature. | Poor air removal or blocked exhaust, solenoid stuck closed or failed coil, kinked exhaust tubing, calibration imbalance between the pressure transducer and the chamber thermocouple; for the low readings also a steam leak or not enough steam. | Check the exhaust tube at the back for kinks and that it is pushed fully into its fitting; on a 2000, clean the venturi at the left rear of the cassette tray. Otherwise bench calibration. |
| CF53 | Rubber and Plastics: the temperature worked out from pressure is more than 2 °C above the measured temperature. | Poor air removal or blocked exhaust, solenoid stuck closed or failed coil, kinked exhaust tubing, calibration imbalance between the pressure transducer and the chamber thermocouple; for the low readings also a steam leak or not enough steam. | Check the exhaust tube at the back for kinks and that it is pushed fully into its fitting; on a 2000, clean the venturi at the left rear of the cassette tray. Otherwise bench calibration. |
| CF54 | Rubber and Plastics: cassette pressure too low for the sterilization temperature (below about 204.8 kPa). | Poor air removal or blocked exhaust, solenoid stuck closed or failed coil, kinked exhaust tubing, calibration imbalance between the pressure transducer and the chamber thermocouple; for the low readings also a steam leak or not enough steam. | Look at the cassette seal (past 500 cycles or 6 months?) and the cassette for warping or dents; replace the seal if in doubt. Check the exhaust tube at the back for kinks and that it is pushed fully into its fitting; on a 2000, clean the venturi at the left rear of the cassette tray. |
| CF55 | Rubber and Plastics: cassette pressure too high (above about 232 kPa). | Poor air removal or blocked exhaust, solenoid stuck closed or failed coil, kinked exhaust tubing, calibration imbalance between the pressure transducer and the chamber thermocouple; for the low readings also a steam leak or not enough steam. | Check the exhaust tube at the back for kinks and that it is pushed fully into its fitting; on a 2000, clean the venturi at the left rear of the cassette tray. |
| CF60 | 134 °C cycles: cassette temperature fell below 134 °C. | Poor air removal or blocked exhaust, solenoid stuck closed or failed coil, kinked exhaust tubing, calibration imbalance between the pressure transducer and the chamber thermocouple; for the low readings also a steam leak or not enough steam. | Look at the cassette seal (past 500 cycles or 6 months?) and the cassette for warping or dents; replace the seal if in doubt. Check the exhaust tube at the back for kinks and that it is pushed fully into its fitting; on a 2000, clean the venturi at the left rear of the cassette tray. |
| CF61 | 134 °C cycles: cassette temperature rose above 138 °C. | Poor air removal or blocked exhaust, solenoid stuck closed or failed coil, kinked exhaust tubing, calibration imbalance between the pressure transducer and the chamber thermocouple; for the low readings also a steam leak or not enough steam. | Check the exhaust tube at the back for kinks and that it is pushed fully into its fitting; on a 2000, clean the venturi at the left rear of the cassette tray. |
| CF62 | 134 °C cycles: the temperature worked out from pressure is more than 2 °C below the measured temperature. | Poor air removal or blocked exhaust, solenoid stuck closed or failed coil, kinked exhaust tubing, calibration imbalance between the pressure transducer and the chamber thermocouple; for the low readings also a steam leak or not enough steam. | Check the exhaust tube at the back for kinks and that it is pushed fully into its fitting; on a 2000, clean the venturi at the left rear of the cassette tray. Otherwise bench calibration. |
| CF63 | 134 °C cycles: the temperature worked out from pressure is more than 2 °C above the measured temperature. | Poor air removal or blocked exhaust, solenoid stuck closed or failed coil, kinked exhaust tubing, calibration imbalance between the pressure transducer and the chamber thermocouple; for the low readings also a steam leak or not enough steam. | Check the exhaust tube at the back for kinks and that it is pushed fully into its fitting; on a 2000, clean the venturi at the left rear of the cassette tray. Otherwise bench calibration. |
| CF64 | 134 °C cycles: cassette pressure too low (below about 304 kPa). | Poor air removal or blocked exhaust, solenoid stuck closed or failed coil, kinked exhaust tubing, calibration imbalance between the pressure transducer and the chamber thermocouple; for the low readings also a steam leak or not enough steam. | Look at the cassette seal (past 500 cycles or 6 months?) and the cassette for warping or dents; replace the seal if in doubt. Check the exhaust tube at the back for kinks and that it is pushed fully into its fitting; on a 2000, clean the venturi at the left rear of the cassette tray. |
| CF65 | 134 °C cycles: cassette pressure too high (above about 341.2 kPa). | Poor air removal or blocked exhaust, solenoid stuck closed or failed coil, kinked exhaust tubing, calibration imbalance between the pressure transducer and the chamber thermocouple; for the low readings also a steam leak or not enough steam. | Check the exhaust tube at the back for kinks and that it is pushed fully into its fitting; on a 2000, clean the venturi at the left rear of the cassette tray. |
Board and software faults: CF70 to CF99
These are about the electronics, not the steam. Several are caused by electrical noise or a static discharge and never come back after a power-cycle; if one repeats, the board needs a bench. Numbers 66–69 are not used.
| Code | What it means | Likely causes | What to do |
|---|---|---|---|
| CF70 | The processor's internal timer and the real-time clock disagree. | A one-off electromagnetic disturbance (will not repeat), or a damaged microprocessor, real-time clock or crystal. | Power-cycle and run again. If it repeats: board repair. |
| CF71 | Pressure reading outside the possible range. | Pressure sensor misconnected, disconnected or damaged; pressure sensor interface board (not on revision 7 boards). | Bench repair |
| CF72 | Microprocessor cannot talk to the Temperature Adapter Board. | Microprocessor or adapter board not seated, bent or broken pins. | Bench repair |
| CF80 | The boiler calibration routine ran at the wrong time. | Electrical noise or static discharge making the microprocessor misbehave. | Power-cycle. If it repeats: board repair. |
| CF81 | The print-data buffer pointer is invalid. | Electrical noise or static discharge making the microprocessor misbehave. | Power-cycle. If it repeats: board repair. |
| CF82 | The calibration-data buffer pointer is invalid. | Electrical noise or static discharge making the microprocessor misbehave. | Power-cycle. If it repeats: board repair. |
| CF83 | The EEPROM signature is corrupted. | Noise corrupting memory or making the processor misbehave. | Bench repair |
| CF84 | A steam generator / chamber calibration value came out of range. | Faulty or miscalibrated temperature measuring circuit, a thermocouple out of tolerance, an old or badly conditioned steam generator, or the calibration routine running at the wrong time. | Bench repair |
| CF85 | Unexpected serial interrupt. | Electrical noise or static discharge making the microprocessor misbehave. | Power-cycle. If it repeats: board repair. |
| CF86 | Unexpected SPI interrupt. | Electrical noise or static discharge making the microprocessor misbehave. | Power-cycle. If it repeats: board repair. |
| CF87 | Unexpected IRQ interrupt. | Electrical noise or static discharge making the microprocessor misbehave. | Power-cycle. If it repeats: board repair. |
| CF88 | Buffer pointers and sample counter set up wrongly. | Electrical noise or static discharge making the microprocessor misbehave. | Power-cycle. If it repeats: board repair. |
| CF89 | The software detected incorrect input parameters. | Electrical noise or static discharge making the microprocessor misbehave. | Power-cycle. If it repeats: board repair. |
| CF90 | Chamber calibration value corrupted or never set. | A new board or microprocessor, or a strong static discharge. | G4: restore NVRAM from the service menu (technician). Otherwise the chamber thermocouple is recalibrated on the bench. |
| CF91 | Pressure calibration value corrupted or never set. | A new board or microprocessor, or a strong static discharge. | Bench: pressure sensor recalibration. |
| CF94 | Controller board memory or display link fault — see the note below on how sources split CF94 and CF95. | In SciCan's chart and the Statim USA G4 sheet as we read them: the board cannot read its 24C512 EEPROM (revision 7 / G4). | Bench: microprocessor and EEPROM, or the revision 7 board, replaced. |
| CF95 | Display link or controller board memory fault — see the note below. | As we read the same sources: the color display controller gets no messages from the revision 7 board — no 5 V on the board (green LED off), a fuse, a cable, or the microprocessor not seated. | Bench: power, fuses and cables checked, then the board. |
| CF96 | G4: the color display controller failed its program check at power-up. | Corrupted display software or a failed display controller. | Bench: software update, or a new display controller. |
| CF98 | The microprocessor cannot talk to the analog-to-digital converter on the board. | Microprocessor not seated, shorted or bent pins, adapter board loose, damaged converter. Power going off and on during a cycle can also show it. | If the power dropped mid-cycle, reset and run again. If it repeats: board repair. |
| CF99 | The microprocessor's watchdog timer tripped. | Software running wrongly, usually from electrical noise or static discharge. | Power-cycle. If it repeats: board repair. |
Messages without a number
| Message | What it means | Likely causes | What to do |
|---|---|---|---|
| Water Quality is Not Acceptable | The water quality sensor in the reservoir reads too many dissolved solids. | Wrong water (tap, filtered, "purified") or a dirty reservoir; less often a sensor or wiring fault. | Drain the reservoir and refill with known-good steam-process distilled water (under about 10 µS/cm). If the message stays, the sensor and board are checked on the bench. |
| Water Level Low, Not Sterile | The unit checks the water level before sterilization; it was low, so the cycle aborted. It beeps for 30 seconds. | Empty or low reservoir; on a G4, a float switch that sticks. | Refill with distilled water, press STOP, rerun. Reprocess the load — it is not sterile. |
| Cycle Interrupted | Either the sterilization phase did not start within 3 minutes of reaching temperature, or the power failed mid-cycle (or was switched off after an error without pressing STOP). | Three in a row of the first kind becomes CF26. A power drop is just a power drop. | Press STOP, rerun the load. If it keeps happening without power problems, treat it as CF26. |
| Press Stop To Reset | Shown with every fault. The unit will not start another cycle until STOP is pressed. | Any fault. | Note the fault first, then press STOP. |
| Cycle Aborted | Printed on the cycle record (followed by NOT STERILE) when STOP was pressed or any fault ended the cycle. | Operator stop or any cycle fault. | Reprocess the load. Look for the cycle fault that caused it. |
| Printer Fault | An optional printer is fitted and not printing. | Paper jam, paper loaded the wrong way, printer switched off, cable, failed printer. | Clear the paper, check the printer switch and that paper feeds from the top of the roll. Otherwise bench. |
| NO CONFIGURATION EEPROM | The microcontroller cannot talk to its configuration EEPROM. | Microprocessor not seated, EEPROM damaged, wrongly inserted or missing, controller board. | Bench repair |
Symptoms that do not show a code
| Symptom | What it means | Likely causes | What to do |
|---|---|---|---|
| GFI trips when the unit is switched on | The outlet's ground-fault breaker trips at power-up. | A leaking check valve. | Bench repair |
| Clicking when the cassette is removed | The solenoid clicks on and off. | Steam leaking from the cassette gets into the cassette microswitch. | Fix the cassette leak (usually the seal). The clicking stops within about a day. |
| Water dripping from the drain tube under the unit | Water under the Statim. | Leaking cassette seal or cassette. | Replace the seal; if it continues, the cassette is checked. |
| Steam from the condenser bottle vent | Steam escaping from the waste bottle. | Water in the bottle below its MIN line. | Keep the condenser bottle filled to the MIN line. |
| Steam at the push-in fitting at the back | Steam escaping where the exhaust tube enters the unit. | Exhaust tube not fully inserted. | Push the tube past the first resistance until it seats. |
| Loud buzzing | A buzzing noise from the unit. | Dirty or failing solenoid valve. | Bench: solenoid cleaned or replaced. |
| Wraps still wet after drying | Wrapped loads come out damp. | Dirty air filters, dirty cassette, a dip or loop in the drain tube, stacked wraps, unit not level, weak compressor airflow. | Replace filters, clean the cassette, run the drain tube straight to the bottle, do not stack wraps, level the unit. If still wet: compressor check. |
When it goes to the bench
Anything that points at the pump, solenoid, check valve, pressure relief valve, thermocouples, calibration or the controller board needs the cover off, a control box or the service menu, and on calibration faults a calibration cassette. That is our work. A seal and cassette check runs from $249, solenoid and valve repairs from $399, a pump from $499 and board work from $449 — all on the pricing page. A Statim 5000 loaner with its own cassette rides along on the pickup for $199, free on repairs over $900. Model specifics are on the Statim 5000 and Statim 2000 pages.
Questions
Can we clear a Statim cycle fault ourselves?
You clear the message by pressing STOP. Whether the fault is gone is another matter: seal, cassette, exhaust tube, water and load faults are fixable at the office; anything that points at the pump, solenoid, check valve, thermocouples, calibration or the board needs the cover off and a bench.
Is the load sterile after a cycle fault?
No. Any cycle that ends on a fault is not sterile, and the printout marks it so. Reprocess the instruments once the unit runs a complete cycle.
Why does the same code have different causes on different Statims?
SciCan standardized the CF numbers from the revision 2.x boards onward, but the causes depend on the model (2000 or 5000, S or non-S) and the software revision. That is why the first thing to note is the revision that flashes in the top right corner of the screen at power-up.
Still not working?